7 Maintenance

40

Melting Point M-560 Operation Manual, Version D

• Remove the cover.
• Follow the instructions mentioned below on the use of the cleaning tool 051978.

• If a melting point capillary is broken and is stuck

in the heating block, use the

cleaning tool

(Order number: 051978).

• Remove the cover and glass holder with glass

window.

• If glass parts of a melting point capillary are

stuck in a hole of the metal heating block, use
the cleaning tool as shown in the picture.