The gauge will require a few basic adjustments in order to get underway with

the calibration process. For this example, in order to avoid multiple steps for

setting the gauge features/options, the default setup will be loaded as a starting

point. However, the user should invest the necessary time getting to know the

features and options included in the DFX-7, as well as gaining a good

understanding of their function:

Loading Default Setup

1) Press the MENU key once to activate the menu items tab. Press the MENU

key multiple times to tab right and the ESC key multiple times to tab left until

the SEUTP menu is highlighted and displaying the submenu items.

2) Use the UP and DOWN arrow keys to scroll through the sub menu items

until DEFAULT SETUP is highlighted.

3) Press the ENTER key to display the CONFIRM screen.

4) Press the OK key to load the DEFAULT SETUP.

Setting the Range

1) Press the MEAS key once to activate the measure menu items. Press the

MEAS key multiple times to move right and the ESC key multiple times to

move left, until the RANGE cell is highlighted.