DFX-7 Ultrasonic Flaw Detector

43

stable. Also, be sure there is sufficient signal amplitude in order to avoid ‘peak

jumping’ or missing the first cycle.

Adjusting the Gate

1) Press the MEAS key once to activate the measure menu items. Press the

MEAS key multiple times to move right and the ESC key multiple times to

move left, until the GATE1 cell is highlighted.

2) Press the LEFT and RIGHT arrow keys to move the gate over the waveform

to be measured, as demonstrated in the screen shot above.

Note: In order to trigger on a signal, the waveform must be inside the

boundaries of the gate.

Calibrating the Second Point

1) Press the MENU key once to activate the menu items tab. Press the MENU

key multiple times to tab right and the ESC key multiple times to tab left until

the CAL menu is highlighted and displaying the submenu items, as shown in

the screenshot above.

2) Use the UP and DOWN arrow keys to scroll through the sub menu items

until MATL 2PT is highlighted.

3) Press the ENTER key to display the Digits Edit Box.