Checkline DFX-7 User Manual

Page 75

Advertising
background image

71

CHAPTER TEN
ADDITIONAL FEATURES OF THE DFX-7

10.1 Material Velocity

The DFX-7 has a built in feature to manually adjust the material velocity to a know
value in the calibration section of the DFX-7. It should be noted that this does not
calculate the zero value like the two point calibration option does. Therefore, if the
material velocity is changed you should check the calibration. The procedures for
adjusting the velocity are outlined below:

Adjusting Velocity

1) Press the MENU key once to activate the menu items tab. Press the MENU

key multiple times to tab right and the ESC key multiple times to tab left until

the CAL menu is highlighted and displaying the submenu items.

2) Use the UP and DOWN arrow keys to scroll through the sub menu items

until VELCITY is highlighted.

3) Press the LEFT and RIGHT arrow keys to scroll the value. When the

correct velocity value is being displayed, proceed to step 8.

4) Alternatively, press the ENTER key to display the Digits Edit Box.

5) Press the UP and DOWN arrow keys to scroll the highlighted value.