DFX-7 Ultrasonic Flaw Detector

37

2) Press the ENTER key to display the Digits Edit Box.

3) Press the UP and DOWN arrow keys to scroll the highlighted value.

4) Press the LEFT and RIGHT arrow keys to scroll the digit locations.

5) Repeat steps 3 & 4 until the desired range value is displayed.

6) Press the OK key to set the value and return to the GT1 menu.

Note: Our range calibration range for this example is 4 inches. Therefore

setting a range of 5 inches will enable the user to view the entire range on the

display.

Enabling and Setting Up a Gate

Refer to the illustrations above. In order to measure, a gate is required. The

width of the gate should be small enough, in terms of its width, in order to select

specific reflections to measure. The screenshot above shows a gate enabled to

select the 1 inch calibration standard, and in the second the 4 inch calibration

standard.