Assette, Loading the cassette – ChemoMetec NC-200 User Manual

Page 21

Advertising
background image

2 Cell counting with the NucleoCounter NC-200

991-0203 Rev. 1.1 Sep. 2012 7

2.2

The Cassette

Figure 2 The Via1-Cassette. Here fluorescent dyes AO (acridine orange) and DAPI (4',6 -

diamidino-2-phenylindole) are immobilized in the first part of the flow system.

2.2.1

Loading the Cassette

The cassette is loaded by gently pressing the white piston down to the top of the cassette,

which creates a partial vacuum in the flow system. The tip of the cassette must be

immersed into a sample mixture when pressing the piston, resulting in sample being

loaded into the flow system. Approximately 60µl is loaded into the flow system filling the

first part of the channel in the cassette. The liquid is transferred to the measurement

chamber when the cassette is inserted in the NucleoCounter NC-200 instrument and an

analysis is initiated. Please note that approximately 100 µl in a normal micro centrifuge

tube is recommended to reduce the risk of air being loaded into the system.

Figure 3 Loading of the Cassette