5 Operation of the NucleoCounter

991-0203 Rev. 1.1 Sep. 2012 14

Figure 9. Illustration of how to hold and place the cassette in the fixture

5.2.1

Actuator movements

When a cassette is analyzed after Run is pressed, the piston actuator will move the piston

rod down through the cylinder of the cassette. Hereby, the sample material is transported

to the cassette measurement chamber. The form and dimensions of the flow channels

facilitate effective mixing of the sample. Sensors monitor sample flow. When the sample

has been correctly loaded into the measurement chamber, the pi ston is stopped.

5.3

Result presentation

Depending on the type of analysis performed the result will be present ed on the PC within

a few seconds and the fluorescent image(s) will be stored on the hard disk. Typical process

time is less than a minute for viability analysis using one Via-1 Cassette.

5.4

Protection of the optical system

The lid covers the fixture. This lid is designed to protect the area from dust and other

potential contaminants, and to keep external light from interfering with the fluorescent

image recording. The lid is mounted with a magnetic hinge.

Unless inserting or removing a cassette or cleaning the fixture, the lid should be closed at

all times to protect the optical system of the NucleoCounter NC-200.