Initial start-up – AERCO BMK 1.5 LN Dual Fuel June 2010 User Manual

Page 39

Advertising
background image

INITIAL START-UP

NATURAL

GAS INLET

PROPANE

INLET

PROPANE

SSOV WITH

REGULATOR

NATURAL

GAS

SSOV WITH

REGULATOR

BRASS HEX HEAD CAP

(REMOVE TO ADJUST

PROPANE REGULATOR)

Figure 4-7

Regulator Adjustment Screw Location

14. Compare the measured oxygen % level to

the range shown below. Also, ensure that
the carbon monoxide (CO) and nitrogen
oxide (NOx) readings do not exceed the
values shown.

Combustion Oxygen Level at

70% Valve Position

Oxygen %

± 0.2

Carbon

Monoxide

5.4 %

<100 ppm

15. If the oxygen level is not within the specified

range, adjust the level using the up (▲) and
down (▼) arrow keys on the VFD. Using the
up (▲) arrow key will increase oxygen level
and the down (▼) arrow key will decrease
the oxygen level.

16. Once the oxygen level is within the specified

range at 70%, lower the valve position to
60%.

NOTE

The remaining combustion calibration
steps are performed using the
Combustion Cal Menu

included in the C-

More Control System. The combustion
calibration control functions will be used
to adjust the oxygen level (%) at valve
positions of 60%, 45%, 30% and 16% as
described in the following steps. These
steps assume that the inlet air
temperature is within the range of 50°F to
100°F.

17. Press the MENU Key on the front panel of

the C-MORE and access the Setup menu.
Enter password 6817 and then press the
ENTER

key.

18. Press the MENU Key on the front panel of

the C-MORE until Combustion Cal Menu
appears on the C-More display.

19. Press the ▲ arrow key until SET Valve

Position

appears on the C-MORE display.

20. Press the CHANGE key. SET Valve Position

will begin to flash.

21. Press the ▲ arrow key until the SET Valve

Position

reads 60%. Press the ENTER key.

22. Next, press the

down (▼) arrow key until

CAL Voltage 60%

is displayed.

23. Press the CHANGE key and observe that

CAL Voltage 60%

is flashing.

24. The oxygen level at the 60% valve position

should be as shown below. Also, ensure that
the carbon monoxide (CO) and nitrogen
oxide (NOx) readings do not exceed the
values shown.

Combustion Oxygen Level at

60% Valve Position

Oxygen %

± 0.2

Carbon

Monoxide

NOx

5.0 %

<50 ppm

<30 ppm

25. If the oxygen level is not within the specified

range, adjust the level using the ▲ and ▼
arrow keys. This will adjust the output
voltage to the blower motor as indicated on
the display. Pressing the ▲ arrow key
increases the oxygen level and pressing the
down ▼ arrow key decreases the oxygen
level.

26. Once the oxygen level is within the specified

range at 60%, press the ENTER key to store
the selected blower output voltage for 60%
valve position.

NOTE

The remaining steps basically repeat the
procedures in steps 19 through 26 for
valve positions of 45%, 30% and 16%.
However, since oxygen levels vary, these
steps are repeated in their entirety. When
performing these steps, also ensure that
the carbon monoxide (CO) and nitrogen
oxide (NOx) readings do not exceed the
values shown for each valve position.

4-7

Advertising