12

cause the housing to deform due to the different

cooling rates of the surfaces.
Do not remove the cover before the thermal barrier

is in place.
Since the heat absorbed by the thermal barrier

continues to affect the temperature of the heat

sink and data logger, remove these parts from the

thermal barrier as soon as the test is completed.

Allow them to cool down before further use. In

most cases cooling overnight is sufficient.

Always take the heat sinks out of the thermal

barrier. Shake the heat sinks to find out whether

the cooling agent has changed form solid to liquid

state. If the cooling agent is still in a solid state, the

heat sinks need to cool down for 8 hours at a tem-

perature lower than +10 °C. If the cooling agent

has remained solid, it is sufficient to cool down the

heat sinks for 8 hours at a temperature lower than

+20 °C.
Make sure that mechanically damaged heat sinks

don’t leak.
Contact BYK-Gardner if the heat sink phase

change material leaks. It is a non-toxic waxlike

substance that dries hard powdery white and

has a slight acidic smell. Wait until it is dry before

scraping off any material that has leaked onto the

surface of the barrier.
The cooling agent is a non dangerous substance

or no dangerous mixture in the sense of the

EUROPEAN UNION guideline 67/548/EWG or

1999/45/EG.
Examine the cooled thermal barrier and heat sinks

for damage before storing in a dry environment to

avoid the ceramic insulation absorbing water.
Check the thermal barrier’s seals and closing me-

chanism, rectify any damage before further use.

Getting started