8. Cleaning and Maintenance

Do not use any acetone for cleaning

the unit! The instrument housing is resistant to a

number of solvents, but cannot be guaranteed to

withstand all chemicals. You should therefore use

a soft, moist cloth for cleaning. For cleaning exces-

sive dirt, use propanol.

Do not attempt to make any repairs your-

self! If a malfunction occurs on your measuring

device, our Customer Service department will be

happy to help you as quickly as possible.
Examination

Always take the heat sinks out of the

thermal barrier. Shake the heat sinks to find out

whether the cooling agent has changed form solid

to liquid state. If the cooling agent is still in a solid

state, the heat sinks need to cool down for 8 hours

at a temperature lower than +10 °C. If the cooling

agent has remained solid, it is sufficient to cool

down the heat sinks for 8 hours at a temperature

lower than +20 °C.

Make sure that mechanically damaged heat

sinks don’t leak.

temp-gard Data Loggers

Store in a dust free environment. Remove

batteries when storing for more than two to three

weeks.
Return to BYK-Gardner annually for calibration.

Thermocouple Probes

General:
Lifting or removing the probes by the cables will

damage the cables and connectors.
Examine cables, and replace any probes found to

have damaged insulation.