Buffalo Filter Turbo Auto-Sense™ Operator’s Manual: Page 22

3.1

General Maintenance Information

This section contains information for ordinary upkeep of the BUFFALO FILTER Turbo Auto-

Sense™ Smoke Evacuation System. While the system has been designed and manufactured to

high industry standards, it is recommended that periodic inspection and performance testing be

performed by a qualified Biomedical Technician to ensure continued safe and effective operation.

3.2 Cleaning

Unplug Unit. Wipe unit with a damp cloth containing mild disinfectant solution or soapy water.

Wipe dry with a clean cloth. Do not steam sterilize.

3.3

Periodic Inspection

The Turbo Auto-Sense™ Smoke Evacuation System should be visually inspected at least every year.

This inspection should include checks for:

• Damage to the power cord.

• Damage to the power plug or power inlet module.

• Proper mating, cleanliness and absence of damage to the filter inlet.

• Obvious external or internal damage to the system.

3.4

Troubleshooting the System – see below.

Maintenance

Section 3.0

PROBLEM POTENTIAL CAUSE CORRECTIVE ACTION

1. Smoke Evacuation System

is ON but suction is minimal

or none.

2. Smoke Evacuation System

does not function even

though suction ON button is

depressed.

1. Filter is not seated

completely.

2. Filter is clogged.

3. Vacuum hose or tube is

clogged.

4. Motor/blower is obstructed.

1. Not plugged into an

electrical outlet.

2. Fuses are blown.

3. Electronic system failure.

4. Filter life has expired or

invalid filter inserted.

1. Re-install ViroSafe® Filter, press firmly into place

and fully engage clip.

2. Replace filter with a genuine ViroSafe® Filter.

3. Replace vacuum hose or tube with genuine

Buffalo Filter® products.

4. Call BioMed or Buffalo Filter Technical Services

at 1.800.343.2324 or +1.716.835.7000.

1. Check power outlet and connection to rear or

side panel of the machine.

2./3. Call BioMed or Buffalo Filter Technical Ser-

vices at 1.800.343.2324 or +1.716.835.7000

4. Replace filter with a genuine Bovie® SF35 filter.