1 introduction, 2 inspection – Buffalo Filter PlumeSafe Turbo User Manual

Page 4

Advertising
background image

Buffalo Filter Turbo Auto-Sense™ Operator’s Manual: Page 3

1.1 Introduction

BUFFALO FILTER® Turbo Auto-Sense™ Smoke Evacuation Systems are intended to evacuate and

filter surgical smoke plume and aerosols created by the interface surgical tools with tissue, exam-

ples being lasers, electrosurgery systems, and ultrasonic devices.

The Turbo Auto-Sense™ Smoke Evacuation Systems have been designed with a high suction, high

flow rate centrifugal action pump. The ultra-quiet motor is used to draw the surgical smoke from

the surgical site through the vacuum tubing and into the ViroSafe® filter where the surgical smoke

is processed by a series of filters. A single disposable filter is used to simplify the installation and

removal during filter changes. The filter is completely enclosed to protect the healthcare person-

nel from potential contamination during filter changes. One Buffalo Filter ViroSafe® filter con-

tains four different stages within to capture the smoke plume.

The first stage filtration is a prefilter whose function is to trap and remove gross particulate and

casual fluid.

The second stage filtration is ULPA grade (Ultra Low Penetration Air) filter whose high-tech

patented (U.S. Patent #5874052) design captures particulates and micro-organisms from .1 to .2

microns at an efficiency of 99.999%.

The third stage filtration uses the highest grade virgin activated carbon, especially designed for

Buffalo Filter for the removal and adsorption of odors and toxic gases produced by burning tis-

sues. These harmful gases may constitute a health hazard to healthcare professionals who are

subjected to prolonged exposure. The activated carbon used in the Turbo Auto-Sense™ Smoke

Evacuation Systems preferentially removes toxic organic gases rather than water vapor and pro-

vides optimal odor removal.

The fourth stage filtration is an expanded foam used to trap activated carbon fines from migrating

out of the filter.

The electronic controls on the face panel of the Turbo Auto-Sense™ Smoke Evacuation System has

been designed “user friendly” and facilitate unit set up and operation. Please refer to Section 2.0

for Operating Instructions.

1.2 Inspection

The Turbo Auto-Sense™ Smoke Evacuator has been thoroughly tested and inspected before ship-

ment from the factory. Please check the unit before using it to insure that no damage has oc-

curred in transit. If damage is evident, please contact BUFFALO FILTER Customer Service at

1-800-343-2324 (US Only) or (716) 835-7000 (International)

System Description

Section 1.0

Advertising