9.2 P1-02 Adjustment test

Here, deviation from the last adjustment is determined. Only a check will be carried
out, no variables will be changes.


Return to weighing mode:

Press the ESC-key repeatedly until the query ”SAVE?“ appears.

Confirm query by pressing the PRINT-key or reject it by pressing the ESC-
key.

Wait until the weighed value for
the required adjustment weight
appears. Place adjustment weight
in the centre of the sample dish.











The results of the adjustment test
will be displayed.