Preparation:

 Switch off instrument.
 Remove sample dish, dish holder, removal

aid and windshield.

 Install the temperature-calibration set acc.

to fig.

 Switch on instrument.

Begin temperature adjustment:

 Close heating hood and carry out

temperature calibration as described in the
following.

 On the temperature calibration set switch-

on the digital thermometer by the ON/OFF
button.