If an optional printer is connected and function “P1-04” is activated, a
adjustment record will be produced.









Confirm input, the third heating phase is

starting.

Temperature calibration of the third point will take 8
min after which you will hear an acoustic signal.
Compare flashing temperature display with the
temperature reading displayed on the
thermometer.

If the two values do not match, correct with the
help of the navigation buttons

 .

confirm, the instrument is returning to the

menu.

Return to weighing mode:

Press the ESC-key repeatedly until the query
SAVE?“ appears.

Confirm query by pressing the PRINT-key or reject
it by pressing the ESC-key.