• P4-01-AuE –Autozero
This function is used to tare small variations in weight automatically.
In the event that small quantities are removed or added to the material to be
weighed, incorrect weighing results can be displayed due to the “stability
compensation“. (e.g. slow flow of liquids from a container placed on the balance,
evaporating processes).
With small weight variations it is thus preferable to switch this function off.
Auto = On
- Function activated
Auto = OFF - Function deactivated
Return to weighing mode:
Press the ESC-key repeatedly until the query ”SAVE?“ appears.
Confirm query by pressing the PRINT-key or reject it by pressing the ESC-
key.