KERN & SOHN ALJ 310-4A User Manual

Page 34

Advertising
background image

ALJ/ALS/PLJ/PLS-BA-e-1338

34

7.2.2 Overwrite internal adjustment weight

(Models with settings not appropriate for verification only)

!

Overwriting is restricted to specialist staff possessing well acquainted with the
workings of weighing scales.

Press the MENU button and keep it pressed until the acoustic signal gets mute.

Press the MENU

button as often as required until „CAlib“ appears

Operate the PRINT key

Press the MENU

button as often as required until „TEC-CAL“ appears

 Press the PRINT button and keep it pressed until the acoustic signal gets mute.

 Press the MENU button and keep it pressed until the acoustic signal gets mute. The

balance changes automatically into weighing mode

 Ensure that there is no load on the weighing plate
 Press the CAL key

Wait until the exact value of the adjustment weight will be shown flashing

(Example)

 Place the displayed adjusting weight on the weighing plate.
The flashing display extinguishes and the balance changes into weighing mode.
 Remove adjustment weight from weighing plate
 Press the PRINT button and keep it pressed until the acoustic signal gets mute.
The calibrating process is started
During this

process „tEc MEM“ is displayed

After having automatically saved the value of the internal adjustment weight, the balance
will return into weighing mode.

 Carry out the adjustment process as described in chapter 7.2

Advertising
This manual is related to the following products: