10.4 Checkweighing

An upper and a lower limit value can be established to ensure that the weighed
goods are exactly within the established tolerance limits.
The display [L], [ok] or [H] shows whether the goods to be weighed are within the
two tolerance limits.

Goods to be weighed below tolerance limit

Goods to be weighed within tolerance range

Goods to be weighed above tolerance limit


The indetermination if the goods to be weighed are within the tolerance limits, can
additionally be given by an acoustic signal.

 In weighing mode press MENU

button. The first menu item „count“ is displayed.

 Press MENU button repeatedly

 Confirm using the PRINT button, the display to enter the lower tolerance limit

appears.

 Enter the lower limit value using the arrow buttons   
 Confirm using the PRINT button, the display to enter the upper tolerance limit

appears.

 Enter the upper limit value using the arrow buttons   

 Confirm using the PRINT button, the display to adjust the signal tone appears.
 Select the desired settings by pressing the MENU button