KERN & SOHN ALJ 310-4A User Manual

Page 6

Advertising
background image

ALJ/ALS/PLJ/PLS-BA-e-1338

6

KERN

ALS 160-4A

ALS 250-4A

Weighing range (max)

160 g

250 g

Readability (d)

0.1 mg

0.1 mg

Reproducibility

0.1 mg

0.1 mg

Linearity

± 0.3 mg

± 0.3 mg

Stabilization time (typical)

4 sec.

4 sec.

Minimum unit weight at
piece counting

> 0.5 mg

> 0.5 mg

Warm-up time

8 hours

Recommended adjustment weight,
not added (class)

150 g (E2)

200 g (E2)

Reference quantities at
piece counting

10, 25, 50, 100 freely selectable

Weighing Units

ct, g, gn, lb, mo, oz, ozt,

tl (HK), tl (Singap. Malays), tl (Tw), pen

Electric Supply

24 V AC, 500 mA

Operating temperature

+ 5° C …. + 35° C

Humidity of air

max. 80 % (not condensing)

Housing (B x D x H) mm

210 x 340 x 330

Dimensions of windshield
(B x D x H) mm

160 x 140 x 205 (inside)

190 x 195 x 225 (outside)

Weighing plate (stainless steel)

Ø 80 mm

Weight kg (net)

5.7 kg

Interface

RS 232C


Advertising
This manual is related to the following products: