KERN

PLJ 310-3F

PLJ 420-3F

PLJ 720-3A

Weighing range (max)

310 g

420 g

720 g

Readability (d)

0.001 g

0.001 g

0.001 g

Reproducibility

0.001 g

0.001 g

0.001 g

Linearity

± 0.003 g

± 0.003 g

± 0.002 g

Stabilization time (typical)

3 sec.

2 sec.

2 sec.

Minimum unit weight at
piece counting

> 5 mg

> 5 mg

> 5 mg

Warm-up time

4 hours

4 hours

4 hours

Adjustment weight

internal

Reference quantities at
piece counting

10, 25, 50, 100 freely selectable

Weighing Units

ct, g, gn, lb, mo, oz, ozt,

tl (HK), tl (Singap. Malays), tl (Tw), pen

Electric Supply

230V/50 Hz

(Euro) 9V DC

230V/50 Hz

(Euro) 9V DC

230V/50 Hz

(Euro) 24V AC

Operating temperature

+ 5° C / + 35° C

Humidity of air

max. 80 % (not condensing)

Underfloor weighing

-

-

Clevis type

eyelet, standard

Housing (B x D x H) mm

210 x 340 x 160

Windshield mm
inside Ø 150, 60 high
outside Ø 160, 70 high

yes

yes

yes

Weighing plate
(stainless steel)

Ø 11 cm

Ø 11 cm

Ø 11 cm

Weight kg (net)

4 kg

4.5 kg

Interface

RS 232C


Advertising