Instant help – KERN & SOHN GAB 30K5DNM User Manual

Page 40

Advertising
background image

40

GAB-N-BA-e-1315

14 Instant help

In case of an error in the program process, briefly turn off the balance and disconnect

from power supply. The weighing process must then be restarted from the beginning.
Help:
Fault

Possible cause

The displayed weight does

not glow.

The balance is not switched on.

Mains power supply interrupted (mains cable

defective).

Power supply interrupted.

Rechargeable battery inserted incorrectly or empty

The displayed weight is

permanently changing

Draught/air movement

Table/floor vibrations

Weighing plate has contact with other objects.

Electromagnetic fields / static charging (choose

different location/switch off interfering device if

possible)

The weighing result is

obviously incorrect

The display of the balance is not at zero

Adjustment is no longer correct.

Great fluctuations in temperature.

Warm-up time was ignored.

Electromagnetic fields / static charging (choose

different location/switch off interfering device if

possible)

Should other error messages occur, switch balance off and then on again. If the error

message remains inform manufacturer.