11 External adjustment "Unverified weighing systems „

(menu block "02 EC”)

• Prepare the required adjustment weight. The weight to be used depends

on the capacity of the scale. Carry out adjustment as near as possible to
the scale’s maximum weight. Info about test weights can be found on
the Internet at: http://www.kern-sohn.com.

• Observe stable environmental conditions. Stabilisation requires a certain

warm-up time.


Procedure:
 In the weighing mode press the BG/NET and ZERO key simultaneously, "01 FnC"

is indicated. Select "02 EC" with the navigation keys.

 Press the TARE button, the currently set adjustment weight is displayed. The last

digit flashes.

To change the adjustment weight select the digit to be changed with BG/ NET
key, the active digit flashes. Enter the desired value with the navigation keys.
Press the TARE key repeatedly, until "-----“ is indicated.