KERN & SOHN KXS-TM User Manual

Page 96

Advertising
background image

Example 4: 5-point linearization/ 30 kg (5kg, 5kg, 10 kg, 10kg)

Display Button

Explication

03 CLn

TARE

Linearization function

L0

TARE

zero point

L11

TARE

First adjustment weight (5 kg)

L21

TARE

Second adjustment weight (5kg), factor 1, i.e., the same size as the first

L32

TARE

Third adjustment weight (10 kg); factor 2, i.e. twice as high as the first

L42

Fourth adjustment weight (10 kg); factor 2, i.e. twice as high as the first

L51

BG/NEt

Terminate linearization (5-point linearization)

03 CLn

Example 4: 7-Point linearization / 30 kg (1 kg, 2 kg, 5 kg, 10 kg, 2 kg, 10 kg)

Display Button

Explication

03 CLn

TARE

Linearization function

L0

TARE

zero point

L11

TARE

First adjustment weight (1 kg)

L22

TARE

Second adjustment weight (2 kg); factor 2, i.e. twice as high as the first

L3 5

TARE

Third adjustment weight (5 kg); Factor 5 i.e. 5 times as high as the first

L4 A

TARE

Forth adjustment weight (10 kg); Factor A i.e. 10 times as high as the first

L52

TARE

Fifth adjustment weight (5kg); factor 2, i.e. twice as high as the first

L6 A

TARE

Sixth adjustment weight (10 kg); Factor A i.e. 10 times as high as the first

L71

BG/NEt

Terminate linearization (7-point linearization)

03 CLn

96

KXS/KXG-TM-BA_IA-e-1420

Advertising