KERN & SOHN KXS-TM User Manual

Page 89

Advertising
background image

Example 015000

15.000kg

 Confirm with the TARE key, the display "000003" appears for the input of

"resolution (d), decimal places and type of scale".

Settings:

Description

(o)

Readability (d), select from 1, 2, or 5 steps

(p)

Number of decimal places, select from 0 ~ 5

0

0

1

0.0

2

0.00

3

0.000

4

0.0000

5

0.00000

(q)

0

Standard weighing unit
Always use setting of "0

.

(r)

0 , 1

⇒ Single-range balance

2

Divide into maximum of two sections (subranges ½ max)

3

Divide into maximum of three sections (subranges 1/6 and 2/3 max)

(s)

1

Order of weighing units in which can be toggled using the UNIT key.
Always use setting of "1

.

(t)

0

⇒ Multi-interval balance

1

⇒ Multi-range balance

 Enter the desired parameters with the navigation keys.