KERN & SOHN KXS-TM User Manual

Page 95

Advertising
background image

Setting "factor”

1 =

All weight values are equal

2 =

The following weight value is twice as high as the first.

3 =

The following weight value is three times higher than the first

4 =

….5 times…

5 =

….6 times…

6 =

….7 times…

7 =

….8 times…

8 =

….9 times…

A =

….10 times…

B =

….11 times…

C =

….12 times…

D =

….13 times…

E =

….14 times…

F =

….15 times…

Examples:

Example 1: 4-point linearization/ 30 kg (10 kg, 10 kg, 10 kg)

Display Button

Explication

03 CLn

TARE

Linearization function

L0

TARE

Zero point

L11

TARE

First adjustment weight (10 kg)

L21

TARE

Second adjustment weight (10 kg)

L31

TARE

Third adjustment weight (10 kg)

L41

BG/NEt

Terminate linearization (4-point linearization)

03 CLn

Example 2: 5-point linearization/ 30 kg (5 kg, 10 kg, 10 kg, 5kg)

Display Button

Explication

03 CLn

TARE

Linearization function

L0

TARE

Zero point

L11

TARE

First adjustment weight (5 kg)

L22

TARE

Second adjustment weight (10 kg); factor 2, i.e. twice as high as the first

L32

TARE

Third adjustment weight (10 kg); factor 2, i.e. twice as high as the first

L41

Fourth adjustment weight (5 kg), factor 1, i.e., the same size as the first

L51

BG/NEt

Terminate linearization (5-point linearization)

03 CLn

KXS/KXG-TM-BA_IA-e-1420

95

Advertising