 Use the navigation buttons to select desired settings

Example unit 20:

Number of balancing units "2

(a), standard weighing unit „kg

(b),

 Confirm with TARE key, the current standard weighing unit is indicated for 2

seconds (e.g. „kg

) followed by the display „000001

for adjustment to the

capacity (max.).

Settings maximum:

Description

(c)

Standard weighing unit

(d-h)

Capacity (max)

(i-n)

Weight value of adjusting weight

 With the navigation keys (see chapter 2.1.1) enter the desired capacity.

Example 015000

15.000kg

 Confirm with TARE key, the display "000002" appears for the input of the weight

value for adjusting weight.

 Enter the desired weight value with the navigation keys.