 Carefully place the adjusting weight into the centre of the balancing plate and

press the TARE key.
After successful adjustment, an acoustic signal sounds.

 Remove the adjustment weight. Press any key, the scale will return to the menu..

 With the navigation keys (see chapter 2.1.1) select "CSP 00".

 Confirm with TARE key.

 To exit the menu, move the adjustment switch into the „LOCK” position, see

chapter 18.