Guidance and manufacturer’s declaration-electromagnetic immunity

The MTS300K100M; MXS300K100M; MPS200K100M; MPS200K100PM; MWS300K100M;
MWS400K100DM, MWS300K-1LM is intended for use in the electromagnetic environment speci-
fied below.
The customer or the user of the MTS300K100M; MXS300K100M; MPS200K100M;
MPS200K100PM; MWS300K100M; MWS400K100DM, MWS300K-1LM should assure that is used
in such and environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment-guidance

Conducted RF
IEC 61000-4-6














Radiated RF IEC
61000-4-3

3 Vrms
150 KHz to 80 MHz














3 V/m 80MHz to
2,5 GHz

3 Vrms














3 V/m

Portable and mobile RF communications
equipment should be used no closer to
any part of the MTS300K100M;
MXS300K100M; MPS200K100M;
MPS200K100PM; MWS300K100M;
MWS400K100DM, MWS300K-1LM

including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.

Recommended separation distance:
d = 1,2 √P
d = 1,2 √P 80MHz to 800 MHz
d = 2,3 √P 800MHz to 2,5 GHz

Where P is the maximum output power
rating of the transmitter in watts (W) ac-
cording to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).

Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey

a

, should be less than the compli-

ance level in each frequency range

b

.


Interference may occur in the vicinity of
equipment marked with the following
symbol:

NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorp-
tion and reflection from structures, objects and people.

a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and

land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which
the MTS300K100M; MXS300K100M; MPS200K100M; MPS200K100PM; MWS300K100M;
MWS400K100DM, MWS300K-1LM

is used exceeds the applicable RF compliance level above, the

MTS300K100M; MXS300K100M; MPS200K100M; MPS200K100PM; MWS300K100M; MWS400K100DM,
MWS300K-1LM

should be observed to verify normal operation. If abnormal performance is observed, ad-

ditional measures my be necessary, such as re-orienting or relocating the MTS300K100M;
MXS300K100M; MPS200K100M; MPS200K100PM; MWS300K100M; MWS400K100DM, MWS300K-
1LM.

b Over the frequency range 150 kHz to 80 MHz, field strengths should be les than 3 V/m.