13 Verification

If a scale is verified, then a verification office or manufacturer puts a verification mark
and one or several seals (seals are damaged during removal) on or in the housing.
Therefore, scale adjusting without a seal loss is not possible.

13.1 Adjusting


Observe stable environmental conditions. The warm-up time (refer to chapter 1) is
required to ensure the scale stabilisation.

Note:
In the case of verified scales, adjusting is locked with the jumper. To carry out adjust-
ing, the jumper is to be set in the adjusting position (central position). (refer to 12.2).

Operation

Display

Turn the scale on using the [ON/OFF] key.

Press and keep the [

0] key pressed for about 3

seconds until the display shows successively the
„SETUP” and „UNIT” symbol.

SETUP

��

UNKT

��

Press the [TARE] key until the „CAL ib” symbol is
displayed.

CAL

��

K

B

Press the [HOLD] key.

CAL U

��

Press the [TARE] key.

The triangle

 must be displayed at the top right side

of the display. If not, press the [TARE] key.

Press the [HOLD] key repeatedly until the „CAL 0”
symbol is displayed.

CAL U

��

CAL O

��

MPS / MTS / MXS / MWS-BA-e-1417