7 printing function, 1 parameters of rs232 interface – KERN & SOHN MWS 400K100DM User Manual

Page 49

Advertising
background image

Printing Pre-Tare memory (refer also to section 8.6):

To do this, press the PRE-Tare key repeatedly until the display will show the „ni” (M)
symbol.
Short pressing the PRINT key will activate printing of the values saved in 5 memo-
ries.

M1 0.0 kg

M2 7.0 kg

M3 10.0 kg

M4 30.0 kg

M5 50.0 kg

9.7 Printing function


To use this function, you need the RS232 interface cable (available as accessories)
which is connected with the round plug at the terminal back.

Note: In medical applications, only the peripheral equipment meeting EN 60601-1
standard can be connected to the interface.

When a scale is in the weighing mode, pressing the PRINT key will result in output of
the particular data, presented below, via the interface. It is the standard way to out-
put data, which cannot be changed.

G

88.8 kg

Gross weight

T

2.0 kg

Taring weight

N

86.8 kg

Net weight

180.0 cm

Size of patient

24.4 BMI

BMI value

9.7.1 Parameters of RS232 interface

Set parameters of the scale interface on the connected device. It is not possible to
change the scale parameters.
Baud rate:

9600 bps

Parity check:

no

Data length:

8 bits

Stop bit:

1 bit

Handshake:

no or Xon/Xoff

Data code:

ASCII

MPS / MTS / MXS / MWS-BA-e-1417