Recommended separation distance between
portable and mobile RF communications equipment and the MTS300K100M; MXS300K100M;
MPS200K100M; MPS200K100PM; MWS300K100M; MWS400K100DM, MWS300K-1LM
The MTS300K100M; MXS300K100M; MPS200K100M; MPS200K100PM; MWS300K100M; MWS400K100DM,
MWS300K-1LM
is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the MTS300K100M; MXS300K100M; MPS200K100M; MPS200K100PM;
MWS300K100M; MWS400K100DM,
MWS300K-1LM
can help prevent electromagnetic interference by maintain-
ing a minimum distance between portable and mobile RF communications equipment (transmitters) and the
MTS300K100M; MXS300K100M; MPS200K100M; MPS200K100PM; MWS300K100M; MWS400K100DM,
MWS300K-1LM
as recommended below, according to the maximum output power of the communications equip-
ment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
d =1,2√P
80 MHz to 800 MHz
d =1,2√P
800 MHz to 2,5 GHz
d =2,3√P
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where p is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
MPS / MTS / MXS / MWS-BA-e-1417