12 Troubleshooting


The scale should be switched off for a short time following an interruption in the pro-
gram sequence and disconnected from mains supply. It is then necessary to repeat
the weighing process from the beginning.

Interruption

Possible cause

Weight display is not illumi-
nated.

• The scale is not switched on.

• The mains supply connection has been interrupted

(mains cable not plugged in/faulty).

• Check the fuse of the mains adapter / green LED

is illuminated next to the fuse.

• Mains failure.

• Batteries are incorrectly inserted or discharged.

• No batteries.

Weight display changes
continuously

• Draught/air movement
• Table/floor vibrations
• The weighing plate is in contact with foreign

matters or is installed incorrectly.

• Electromagnetic fields/static charging (choose a

different location for the scale, switch off an inter-
fering device if possible).

The weighing result is
obviously incorrect

• The scale display is not set to zero
• Incorrect adjustment.
• Great fluctuations in temperature.
• Warm-up time was ignored.
• Electromagnetic fields/static charging (choose a

different location for the scale, switch off an inter-
fering device if possible).


Should other error messages occur, switch the scale off and then on again. If the
error message remains, inform the manufacturer.