Ultrasonic milkanalyser

Operation manual

04.11.09

16/107

Make sure the analyser is still during measurement. Vibrations may lead to
inaccurate results.


To start measurement:
- pour the preliminary prepared sample in the sample holder (now you may
use holders with different size - plastic and glass 100 ml);

Fig.6 Input pipe









The input pipe of the analyser is knee-joint (mobile). Move it forward in order
to be possible to dip its end in the milk sample and place the sample holder in
the recess of the analyser..

- press the button Enter.
The analyser sucks the milk, makes the measurement and used milk sample
is poured through the holes of the grid below the input pipe of the analyser
and enters the vessel for collecting waste liquids (see Fig. 3 Principle working
scheme of the analyser). During the measurement the temperature of the
sample is shown on the display.
Ignore the results received immediately after switching on the analyser and
after measuring distilled water. Make a second measurement with new
portion of the same sample.

3.2.3. Displaying the results

3.2.3.1. When the measurement is finished, the sample returns in the
sample-holder and the display shows the results. For example:


Results:
F=ff.ff S=ss.ss
D=dd.dd P=pp.pp
L=ll.ll W=ww.ww

Where: