LABEC M-MIA-LA User Manual

Page 47

Advertising
background image

Ultrasonic milkanalyser

Operation manual

04.11.09

45/107


7.2.3.7.4. Set Calibr Name.
Sets the names of the separate calibrations. The name could be chosen from
the group of predefined calibrations names or to edit a new one. When editing
the new name there is a possibility all ASCII codes to be used, as letters (caps
and normal), numbers and punctuation marks and popular symbols. The
calibration name consists of 8 symbols.

Example:
When it is suitable to us this possibility of the analyser? For example if you have
a device factory calibrated for Cow milk, Sheep Milk and UHT milk, but yоu need
oftenly to measure camel milk. Using the methods, explained in details in
Appendix Methods you may make a new calibration without need to send the
analyser back to the producer for calibration. Using this procedure you may
make calibrations for most oftenly analysed milk and to write down the exact
calibration name, which will be shown on the display and printed on the printer.

After starting this menu the display shows:

Select Calibr

Cal1: Sheep

Exit Yes Next

There are the following possibilities:
With button Exit – to leave the menu.
With button Yes – to confirm the chosen for editing calibration name.
With button Next – to choose the next calibration name for editing.
If a calibration for change or edit of name is chosen, the display shows
(example):

Cal:Sheep

PreDef Exit Edit

There are the following possibilities:
With button PreDef – to choose a calibration name from the list of preliminary
given names.
With button Exit – to leave the menu.
With button Edit – to edit the new calibration name.