LABEC M-MIA-LA User Manual

Page 39

Advertising
background image

Ultrasonic milkanalyser

Operation manual

04.11.09

37/107

Recalibrate 1

Put sample High

Cal meas=1/5


Follow the procedure till the 5th measurement.
After 5th measurement completion automatically appears the menu:


Cal 1 Samp Low
Fat = f.ff
SNF=y.yy
Den=d.dd
------------------------------
Lac=l.ll
Sol=s.ss
Pro=p.pp
Exit


Which reminds to enter the values for the sample with low FAT? Enter the
values analogically to the procedure, described in the previous sample. After
entering the last parameter position the cursor on “Exit”, press “Enter” and
the following menu will appear:

Recal 1

Put sample Low 5

times


Make 5 times measurement of the low FAT sample.
After 5th measurement completion automatically appears the menu:

Recal 1

Put sample water


Which reminds for 5-times water measurement? There is no need to enter
values after placing the sample-holder. Measurement starts directly.