VacuuHandControl VHC

The instrument allows the aspiration of excess liquids. The VHC is supplied completely with three differ-

ent adapters (included in delivery) to hold with pasteur pipettes, capillaries and pipette tips (100 µl).

Push the locking ring on the control button (release aspiration

tube).

Switch off pump.

Aspiration of liquid:

Press control button until the liquid has been aspirated. Re-

lease the control button to stop aspiration.

For controlled aspiration of excess liquids, the aspiration

speed can be regulated by using the integrated control but-

ton. Small amounts of liquids can be taken up manually by

using the thumb wheel and then sucked into the collecting

bottle.

Before aspiration:

Push back the locking ring from the control button.

Choose the appropriate adapter and carefully insert the pi-

pette into the adapter.

Position pipette tip by turning the fixing clip in such a way that

working is comfortable and fatigue-proof. Remove adapter

from fixing clip if is hard to turn. Squeeze fixing clip and turn

to desired position. Put adapter back in fixing clip.

Attention: Never insert filled pipettes due to risk of contami-

nation and infection. Wear protective gloves!

Working at different work places:

Disconnect bottle VTC with VacuuHandControl VHC from the

inlet of the pump. The bottle VTC is portable. If the aspira-

tion effect ceases, connect VTC to the diaphragm pump for

a short time until the pump is switched off by the automatic

pressure control (BVC 21 NT) or until the pump reduces its

speed down to standstill (BVC 21 NT VARIO).

After aspiration:

Carefully pull pipette out of the adapter.

Attention: Risk of contamination and infection. Wear protec-

tive gloves and eye protection! Dispose of pipettes according

to appropriate regulations.