Technical data – VACUUBRAND BVC 21 NT VARIO User Manual

Page 9

Advertising
background image

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999157 / 10/12/2009

page 9 of 31

Technical data

We reserve the right for technical modifications without prior notice!

Type

BVC 21 NT VARIO BVC 21 NT

BVC 01

Pump

Pump

Valve

Maximum pumping speed

(ISO 21360)

m

3

/h

2.2

1.9/2.2*

-

Ultimate vacuum (absolute)

mbar

100

150

-

Maximum permissible inlet pres-

sure (absolute)

bar

1.1

Maximum permissible outlet

pressure (absolute)

bar

1.1

Rated power

kW

0.09

0.12

0.03

No-load speed

min

-1

0-1800

1500/1800*

-

Control range

mbar

100 - atmosphere

-

-

Permissible ambient temperature

storage / operation

°C

-10 to +60 / +10 to +40

Permissible relative atmospheric

moisture during operation

(no condensation)

%

30 to 85

Maximum permissible range of

supply voltage

Attention: Observe specifica-

tions of rating plate!

100-120 V~ +5%/-10% 50/60 Hz

230 V~

±10% 50/60 Hz

Rated current at:

100-120 V~ 50/60 Hz

230 V~ 50/60 Hz

A

A

2 (maximum)

0.9 (maximum)

2.6 / 2.6

1.1 / 1.1

-

-

Internal fuse

4A SPT 5x20

-

-

Motor protection

thermal cutout, manual reset

-

Degree of protection IEC 529

IP 20

IP 20

-

System

System

System

Volume of collecting bottle

4

4

4

Adapter for VHC

pasteur pipettes, capillaries, pipette tips (100 µl)

Outlet

hose nozzle DN 10 mm

hose nozzle

DN 6/10 mm

Hydrophobic filter

Type

Manufacturer

Nominal pore diameter

Midisart

®

2000

Sartorius

0.2

Degree of protection IEC 529

IP 20

IP 20

IP 52

Dimensions L x W x H approx.

mm

275 x 450 x 520

275 x 450 x

520

300 x 255 x

500

Weight approx.

kg

14.5

12.2

4

* at 50/60 Hz

Advertising
This manual is related to the following products: