Establish hose connections to the pump head in three

places.

Push transparent PVC hoses (2x) onto the hose connec-

tors (1)+(2).

Push PTFE tubing onto the connector at the quick cou-

pling (3). Fix the PTFE tubing with the clip to the hose

connector. Close clip with flat pliers.

Assembling the connecting hoses

Reassemble in reverse order.

Install head cover, valves and housing cover with housing

cover insert.

Make sure that the

valves are correctly seated: Valves at

the outlet with round centred opening under valve, valves

at the inlet with kidney-shaped opening beside valve.

By turning eccentric bushing, bring connecting rod into up-

per turning point position (Maximum stroke of the rod).

Pay attention that the diaphragm stays positioned centrally

so that it will become clamped uniformly between housing

and head cover.

Screw in four socket head screws fixing housing cover

crosswise first slightly then tighten.

Do not tighten until head cover is in contact with housing,

torque

12 Nm.

Assembling the pump head

By turning eccentric bushing (front of connecting rod),

bring connecting rod into a position in which diaphragm is

in contact with housing and centred with respect to bore.

Tilt pump carefully onto the left side so that the diaphragm

is on top. Support pump appropriately.