Removing cover hood from base plate, Cleaning and inspecting the pump head – VACUUBRAND BVC 21 NT VARIO User Manual

Page 21

Advertising
background image

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999157 / 10/12/2009

page 21 of 31

Use a Torx screw driver TX 20 to remove the 4 securing

screws of the cover hood.

Put device back to normal operating position.

Use a hex key size 2.5 to remove the top 4 securing screws

(see small figure) of the cover hood on the rear side.

Remove cover hood carefully upwards.

Do not cant the cover at the ducts of the quick coupling or

the adjusting knob.

Cleaning and inspecting the pump head

Switch device off and unplug mains cable.

Uncouple hose between collecting bottle and pump at the

quick coupling. Remove bottle from its holder.

BVC 21 NT VARIO:

Remove adjusting knob of vacuum control carefully.

Tilt device carefully onto the left side.

remove

adjusting

knob

BVC 21 NT VARIO:

Loosen the two socket head screws holding the variable

speed drive slightly (half a turn).

Attention: Don’t undo the socket head screws completely

on any account!

Move the variable speed drive in the slots of the base plate

to the back until the axis of the adjusting knob has disap-

peared completely inside the cover hood.

Tighten the two socket head screws again.

Undo hose connections to the pump head in three plac-

es.

Remove transparent PVC hoses (2x) from the hose con-

nectors (1)+(2).

Open the clip at the connection of the PTFE tubing at the

quick coupling (3) with a flat bladed screw driver. Remove

PTFE tubing from hose connector.

Removing cover hood from base plate

1

3

2

Advertising