Write commands, Read command – VACUUBRAND PC 600 LAN User Manual

Page 29

Advertising
background image

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. Manual-no.: 999084 / 05/10/2009

page 29 of 46

Status of

IN_STAT_1 XXXXX

process control

Mode

: see IN_STAT

Status of operation : see IN_STAT
Isolation valve

0: closed
1: open

Cooling water

0: closed

valve

1: open

Relay

0: not operated
1: operated

Write commands

Function

Command

Parameter

Description

Preset pressure

OUT_SP_1

XXXX

In unit according to selection at controller
(0001 to 1060 mbar (hPa) or 0001 to 0795 Torr)

Pumping speed

OUT_SP_2

XX.X

In Hz (01.0 to 60.0, 99.9 corresponds HI)

Start-up pressure/ OUT_SP_3

XXXX

In unit according to selection at controller

upper pressure value

(see OUT_SP_1)

VACUU•LAN: start-up pressure ”p On”
Vacuum control: ”p set” + ”

Δp”

Switch-off time

OUT_SP_4

hh:mm

Hours:minutes

Lower pressure

OUT_SP_5

XXXX

In unit according to selection at controller

value

(see OUT_SP_1)
Vacuum control: upper pressure
Continuous pumping: lower pressure

Time delay

OUT_SP_6

hh:mm

Hours:minutes

(only VACUU•LAN;
otherwise process time)

Mode of operation OUT_MODE

X

0: VACUU•LAN

0S: VACUU•LAN blocked

1: Continuous pumping
2: Vacuum control
4: RC /RC 6 management

Read command

Function

Command Response

Description