Montage of the connecting hose:

Individual performance check of a pump head:

Assembling pump heads

➨ By turning eccentric bushing (front of connecting rod), bring

connecting rod into a position in which diaphragm is in con-
tact with housing and centred with respect to bore.

Reassemble in reverse order.
➨ Install head cover, valves and housing cover with housing

cover insert.

☞ Make sure that the valves are correctly seated: Valves at

the outlet with round centred opening under valve, valves at
the inlet with kidney-shaped opening beside valve.

➨ By turning eccentric bushing, bring connecting rod into up-

per turning point position (Max. stroke of the rod).

☞ Pay attention that the diaphragm stays positioned centrally

so that it will become clamped uniformly between housing
and head cover.

➨ Screw in four socket head screws fixing housing cover cross-

wise (e. g. in the sequence

➀, ➁, ➂, ➃) first slightly, then

tighten.

☞ Do not tighten until head cover is in contact with housing,

torque 12 Nm.

By measuring the pressure at the inlet port of the individual
head:
Use a suitable vacuum gauge (e. g. DVR 2, cat. no.: 68 29 02),
make sure that it is correctly calibrated, and measure the pres-
sure at the inlet port. A vacuum of less than 120 mbar should
be indicated.
☞ If the reading is higher, recheck the pump chamber and make

sure that the valves and the diaphragms are correctly seated
(diaphragms concentric with bore).

➨ Use open ended wrench (w/f 14) to reconnect hose to elbow

fitting.