Write commands – VACUUBRAND PC 600 LAN User Manual

Page 30

Advertising
background image

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. Manual-no.: 999084 / 05/10/2009

page 30 of 46

Connector assignment

2: RxD
3: TxD
4: DTR (12 V)
5: Mass

0: Remote off
1: Remote on
0: Unit of pressure mbar
1: Unit of pressure Torr
2: Unit of pressure hPa
0: External error off
1: External error on (evaluation)
2: Stop (control is stopped)
0: Acoustic signal not configured
1: Acoustic signal configured
0: Isolation valve not existing
1: Isolation valve existing
0: Coolant valve not existing
1: Coolant valve existing
0: Valve
1: Pump
2: VARIO

Write commands

Function

Command Parameter

Description

The single abbreviations of a command are separated by underscores (ASCII 5FH).
Command and parameter are separated by ignore characters (ASCII 20H).
The string is terminated with <CR><LF> (ASCII 0DH, ASCII 0AH).

*

If remote operation is selected or deselected, the user has to ensure that no dangerous status of the
system due to the change of the mode of operation can occur and to provide appropriate safety measures,
especially if selecting remote operation interferes with a locally operated active process.

Start process control

START

End process control

STOP

X

1: End process control
2: End process control and store actual
pressure value as preset pressure

Remote operation*

REMOTE

X

0: Set controller to local operation
1: Set controller to remote operation

Store all parameters

STORE

to EEPROM

Configuration

OUT_CFG

XXXXXXX

Control