VACUUBRAND PC 600 LAN User Manual

Page 37

Advertising
background image

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. Manual-no.: 999084 / 05/10/2009

page 37 of 46

➨ Use open-ended wrench (w/f 14) to turn elbow fitting 1/4 of a

turn, remove hose.

☞ Do not remove the elbow fitting from the pump head. Through

reassembly a leak may result.

➨ To check valves use hex key to remove four socket head

screws from pump head and remove upper housing (hous-
ing cover with housing cover insert), head cover and valves.

☞ Never remove parts by using a spiky or sharp-edged tool

(e.g. screw driver), we recommend to use a rubber mallet or
compressed air (to be blown carefully into port).

➨ Remove head cover from housing cover insert and check

valves. Note position of valves and remove.

☞ Replace valves if necessary.

☞ Use petroleum ether or industrial solvent to remove depos-

its. Do not inhale.

head cover

housing cover with
housing cover insert

valve

connecting rod

housing

diaphragm

diaphragm clamping disc
with square head screw

diaphragm support disc

washer

View of the disassembled pump head parts

Advertising
This manual is related to the following products: