Operation – VACUUBRAND VACUU·CONTROL User Manual
Page 49
Operation
49
BA-N°: 999279/17.10.2014
Meaning
Menu
control parameter
Vac control
Parameter – Vac control
Parameter
Meaning
Set vacuum
(mbar)
Adjustment of lower value for two-point control or
precise for VARIO pumps.
Adjustment range: 0 ‑ 1060 (0 = Turbo mode)*
Speed (%)
Speed parameter for pump down;
Adjustment range: 1 ‑ 101 (101 = HI mode)**
Hysteresis
Control parameter for 2-point controllers.
Adjustment
range: 0; 1 ‑ 1060 (0 = AUTO)
Maximum
(mbar)
Adjustment of upper pressure. If this limit is
exceeded, the vacuum pump switches off.
Adjustment range: 0; 1 ‑ 1060 (0 = OFF)
Duration (min)
Default value for process run time. Process time
begins when user presses “Start“ button.
Adjustment range: 0 ‑ 1440 (0 = OFF)
Delay (min)
Delay time for an optional coolant valve. Coolant
valve is closed after Delay time has been reached.
* Turbo mode: auto-adapting vacuum controller for best ultimate vacuum.
** HI mode: optimum speed for the respective pressure, recommended setting.