Operation

49

BA-N°: 999279/17.10.2014

Meaning

Menu

control parameter

Vac control

Parameter – Vac control

Parameter

Meaning

Set vacuum

(mbar)

Adjustment of lower value for two-point control or

precise for VARIO pumps.

Adjustment range: 0 ‑ 1060 (0 = Turbo mode)*

Speed (%)

Speed parameter for pump down;

Adjustment range: 1 ‑ 101 (101 = HI mode)**

Hysteresis

Control parameter for 2-point controllers.

Adjustment

 range: 0; 1 ‑ 1060 (0 = AUTO)

Maximum

(mbar)

Adjustment of upper pressure. If this limit is

exceeded, the vacuum pump switches off.

Adjustment range: 0; 1 ‑ 1060 (0 = OFF)

Duration (min)

Default value for process run time. Process time

begins when user presses “Start“ button.

Adjustment range: 0 ‑ 1440 (0 = OFF)

Delay (min)

Delay time for an optional coolant valve. Coolant

valve is closed after Delay time has been reached.

* Turbo mode: auto-adapting vacuum controller for best ultimate vacuum.
** HI mode: optimum speed for the respective pressure, recommended setting.