Operation – VACUUBRAND VACUU·CONTROL User Manual
Page 51
Operation
51
BA-N°: 999279/17.10.2014
Menu
control parameter
Program
Meaning
Menu
control parameter
VACUU•LAN
Meaning
Parameter – Program
Parameter
Meaning
Current program
Edit program, up to 10 programs are
storable for CVC 3000.
... load from prog.
Load the selected program.
... save as prog.
Save the program under the selected number.
Parameter – VACUU•LAN
Parameter
Meaning
Set vacuum
(mbar)
Adjustment of the vacuum set point.
Adjustment range: 1 ‑ 1060
Switch on
(mbar)
Maximum limit for vacuum level. If pressure
exceeds this limit, the pump begins to pump down.
Adjustment range: 0 ‑ 1060
Delay (min)
Period for which the pump will attempt to
maintain “Set vacuum” level. If no pressure
increases are detected before the time period
“Delay” expires, vacuum level will be allowed
to rise to “Switch on” level.
Adjustment range: 0 - 300 minutes