Operation
50
BA-N°: 999279/17.10.2014
Meaning
Parameter – Auto mode
Parameter
Meaning
Sensitivity
Motor speed responsiveness setting:
‑ Low:
fast; large amounts of solvents
- Normal: normal
- High:
slow; small amounts, for solvents
with tendency to foam
Speed (%)
For VARIO pumps: Motor speed parameter for
pump down;
For CVC 3000 Detect: Hysteresis parameter
Adjustment range: 1 ‑ 101 (101 = HI mode)*
Minimum (mbar)
For VARIO pumps only: Vacuum setting. Once
reached, controller switches off the VARIO
pump.
Adjustment range: 0; 1 ‑ 1060 (0 = OFF)
Duration (min)
Process runtime. Measured from when user
presses “Start” button. Pump stops when
Duration has been reached.
Adjustment range: 0 ‑ 1440 (0 = OFF)
Delay (min)
Delay time for an optional coolant valve. Coolant
valve is closed after Delay time has been
reached.
* HI mode: optimum speed for the respective pressure, recommended setting.
Menu
control parameter
Auto mode