Operation – VACUUBRAND VACUU·CONTROL User Manual

Page 50

Advertising
background image

Operation

50

BA-N°: 999279/17.10.2014

Meaning

Parameter – Auto mode

Parameter

Meaning

Sensitivity

Motor speed responsiveness setting:

‑ Low: 

fast; large amounts of solvents

- Normal: normal

- High:

slow; small amounts, for solvents

with tendency to foam

Speed (%)

For VARIO pumps: Motor speed parameter for 

pump down;

For CVC 3000 Detect: Hysteresis parameter

Adjustment range: 1 ‑ 101 (101 = HI mode)*

Minimum (mbar)

For VARIO pumps only: Vacuum setting. Once 

reached, controller switches off the VARIO 

pump.

Adjustment range: 0; 1 ‑ 1060 (0 = OFF)

Duration (min)

Process runtime. Measured from when user

presses “Start” button. Pump stops when 

Duration has been reached.

Adjustment range: 0 ‑ 1440 (0 = OFF)

Delay (min)

Delay time for an optional coolant valve. Coolant

valve is closed after Delay time has been

reached.

* HI mode: optimum speed for the respective pressure, recommended setting.

Menu

control parameter

Auto mode