Leaning, Desinfecting – ADLINK MLC 4-21 User Manual

Page 26

Advertising
background image

MLC 4-xx

Technical Manual Rev. 1.2

Cleaning & desinfecting

Document: Manual_Medical_MLC_1.2

Page 26 of 37 pages

Last change date: 18-Feb-15

10. C

LEANING

&

DESINFECTING

PENTA recommends product cleaning and disinfecting according to the guidelines and specifications

of the German Robert Koch-Institute (RKI). These guidelines apply to the MLC only. For cables, sen-

sors and other peripheral items, follow the cleaning instructions in the Directions for Use document

that accompany these accessories.

The following germicidal products are PENTA approved disinfectants for use on your PENTA product,

especially for the touchscreen or glass front:

Incidin plus

Skinmansoft

Cutasept G. & F.

Sterilium

Aseptoman

Use a spray applicator, apply any of the approved liquids from the list and use a soft lint free cloth to

clean the screen. Follow the cleaning agent manufacturer guidelines.

While there are many other options for disinfecting PENTA products, PENTA may not have tested the

effectiveness of these products on your PENTA product. If you use a product or method other than

those that we’ve recommend above, you should document accordingly in your cleaning protocol.

Please contact your local PENTA partner for the approval of further germicides and detergents.

Note: Disinfect according to your facility's procedures and local regulations.

As a general safety precaution, your PENTA product should undergo decontamination before return-

ing the device to PENTA for service, repair, inspection or disposal.

Note: Contaminated items must not be returned without prior, written agreement!

Electric shock hazard. Before cleaning the MLC, disconnect the power cord to

the device.

!