The MLC 4-xx is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an
environment.

Immunity Test

IEC/EN 60601-1-2:2007
(3

rd

edition) Test Level

IEC 60601-1-2:2014
(4

th

edition) Test Level

Device complies to

Electromagnetic Environment - Guid-
ance

Short interruptions on power
supply input lines
IEC 61000-4-11

Dropout to 5% for 10ms

Interrupt > 95% for 5s

Dropout to 0% for 0.5 cycles @
0°, 45°, 90°, 135° , 180°, 225°,
270° & 315° phase angles

Interrupt 100% for 250/300 cy-
cles

IEC/EN 60601-1-2:2007

IEC 60601-1-2:2014
Dropout criterion A
Interrupt 5s criterion C

Mains power quality should be that of a
typical commercial or hospital environ-
ment. If the user of the device requires
continued operation during power mains
interruptions, it is recommended that the
device is powered from an uninterruptible
power supply or a battery.