13.1.3.

Guidance and manufacturer’s declaration - electromagnetic immunity - for EQUIPMENT and SYSTEM that are not LIFE-

SUPPORTING (IEC 60601-1-2:2007 Table 4)

The MLC 4-xx is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an
environment.

Immunity Test

IEC/EN 60601-1-2:2007

(3

rd

edition) Test Level

IEC 60601-1-2:2014

(4

th

edition) Test Level

Device complies to

Electromagnetic Environment - Guid-
ance

Radiated RF

IEC 61000-4-3

3V/m, 80-2500MHz, 1kHz/2Hz
80%AM modulation

10V/m, 80-2500MHz, 1kHz/2Hz
80%AM modulation for life sup-
port equipment

Dwell Time is 3 seconds for 2Hz
modulation

Dwell Time is 1 second for 1kHz
modulation

3V/m, 80-2700MHz, 1kHz
80%AM modulation for Profes-
sional Healthcare Facility Envi-
ronment

10V/m, 80-2700MHz, 1kHz
80%AM modulation for Home
Healthcare Environment

IEC/EN 60601-1-2:2007

Life support equipment

IEC 60601-1-2:2014

Home Healthcare Environment

Portable and mobile RF communications
equipment should be used no closer to
any part of the device, including cables,
than the recommended separation dis-
tance calculated from the equation appli-
cable to the frequency of the transmitter.

Recommended separation distance:

𝑑 = 1.2 Γ— βˆšπ‘ƒ

from 150kHz to 80MHz

𝑑 = 1.2 Γ— βˆšπ‘ƒ

from 80MHz to 800MHz

𝑑 = 2.3 Γ— βˆšπ‘ƒ

from

800MHz to 2.5GHz

where P is the maximum output power
rating of the transmitter in watts (W) ac-
cording to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).

Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey [a], should be less than the compli-
ance level in each frequency range [b].

Interference may occur in the vicinity of
equipment marked with the following sym-
bol:

Conducted RF

IEC 61000-4-6

3Vrms, 0.15-80 MHz, 1kHz/2Hz
80% AM modulation

3Vrms, 0.15-80 MHz, 1kHz/2Hz
80% AM modulation for life sup-
port equipment

3Vrms, 0.15-80 MHz, 1kHz 80%
AM modulation

6Vrms in ISM bands (I/O cables
< 3m excluded)

10Vrms

IEC/EN 60601-1-2:2007

IEC 60601-1-2:2014

NOTE 1 - At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 - These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.