MLC 4-xx

Technical Manual Rev. 1.2

Appendix

Document: Manual_Medical_MLC_1.2

Page 34 of 37 pages

Last change date: 18-Feb-15

13.1.2. Guidance and

manufacturer’s declaration – electromagnetic immunity – for all EQUIPMENT and SYSTEMS (IEC 60601-1-

2:2007 Table 2)

The MLC 4-xx is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an
environment.

Immunity Test

IEC/EN 60601-1-2:2007
(3

rd

edition) Test Level

IEC 60601-1-2:2014
(4

th

edition) Test Level

Device complies to

Electromagnetic Environment - Guid-
ance

Electrostatic discharge (ESD)
IEC 61000-4-2

±6kV contact
±8kV air

±8kV contact
±15kV air

IEC/EN 60601-1-2:2007

IEC 60601-1-2:2014
Contact discharge criterion C
Air discharge criterion A

Floors should be wood, concrete, or ce-
ramic tile. If floors are covered with syn-
thetic material, the relative humidity should
be at least 30%.

Electrical fast transients / Burst
IEC 61000-4-4

±2kV on power supply lines,
5kHz repetition
±1kV for input/output lines

±2kV on power supply lines,
100kHz repetition
±1kV for input/output lines,
100kHz repetition

IEC/EN 60601-1-2:2007

IEC 60601-1-2:2014

Mains power quality should be that of a
typical commercial or hospital environ-
ment.

Surge
IEC 61000-4-5

±1kV differential on AC power
supply lines
±2kV common mode

±(500V, 1kV & 2kV) differential
on AC power supply lines > 3m
±(500V, 1kV & 2kV) common
mode
±2kV common mode (outdoor
cables)

IEC/EN 60601-1-2:2007

IEC 60601-1-2:2014

Mains power quality should be that of a
typical commercial or hospital environ-
ment.

Power frequency magnetic field
IEC 61000-4-8

50 and 60Hz
3A/m

50 and 60Hz
30A/m

IEC/EN 60601-1-2:2007

IEC 60601-1-2:2014

Power frequency magnetic fields should
be at levels characteristic of a typical loca-
tion in a typical commercial or hospital
environment.

Voltage dips and voltage varia-
tions on power supply input lines
IEC 61000-4-11

Dip to 40% for 5 cycles (100ms)

Dip to 70% for 25 cycles (500ms)

Dip to 0% for 1 cycles @
0°phase angle

Dip to 70% for 25/30 cycles @ 0°
phase angle

IEC/EN 60601-1-2:2007

IEC 60601-1-2:2014

Mains power quality should be that of a
typical commercial or hospital environ-
ment. If the user of the device requires
continued operation during power mains
interruptions, it is recommended that the
device is powered from an uninterruptible
power supply or a battery.